Tektronix TDS 340A 100 MHz
"Tektronix TDS 340A characteristics"
by RS232 (2058 pt)
2017-Oct-02 18:58

 Tektronix TDS 340A characteristics

   Signal acquisition system

Sec/Div setting : 10 s/div – 5 ns/div
Minimum pulse width : 10 ns
Measurement type : Average of ≥16 waveforms
DC accuracy : ±(2.0% × |reading| + 0.15 div + 0.3 mV)

Accuracy, Offset - Volts/Div setting :
2 mV/div – 99.5 mV/div
100 mV/div – 995 mV/div
1 V/div – 10 V/div

Analog Bandwidth, DC Coupled : DC – ≥100 MHz

Input Impedance, DC-Coupled : 1 M ±1% in parallel with 20 pF ±2.0 pF

Input Voltage, Maximum : ±300 V (DC or AC) CAT II; derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz and above

   Time base system

Accuracy, Long Term Sample Rate and Delay Time : ±100 ppm over any ≥1 ms interval

   Triggering system

Accuracy, Trigger Level, DC Coupled : ±(3% of |Setting – Net Offset1| + 0.2 div × volts/div setting + Offset Accuracy

Sensitivity, Edge-Type Trigger DC Coupled :
CH1 or CH2 0.35 division from DC to 20 MHz, increasing to 1 div at 100 MHz
External : 50 mV from DC to 20 MHz, increasing to 150 mV at 100 MHz
External/10 :  500 mV from DC to 20 MHz, increasing to 1.5 V at 100 MHz

Input Impedance, External Trigger : 1 M ±2% in parallel with 20 pF ±2 pF

   Power requirements

Power Consumption : ≤65 Watts (120 VA)

   Signal acquisition system

Accuracy, DC Gain, Envelope Acquisition Mode : ±2% for sec/div settings from 10 s/div to 5 ns/div

Accuracy, DC Voltage Measurement Sample Acquisition Mode : DC accuracy ±(2.0% × (|reading – Net Offset1|) + Offset
Accuracy + 0.13 div + 0.6 mV)