"Tektronix TDS 340A characteristics" by RS232 (2058 pt) | 2017-Oct-02 18:58 |
Tektronix TDS 340A characteristics
Signal acquisition system
Sec/Div setting : 10 s/div – 5 ns/div
Minimum pulse width : 10 ns
Measurement type : Average of ≥16 waveforms
DC accuracy : ±(2.0% × |reading| + 0.15 div + 0.3 mV)
Accuracy, Offset - Volts/Div setting :
2 mV/div – 99.5 mV/div
100 mV/div – 995 mV/div
1 V/div – 10 V/div
Analog Bandwidth, DC Coupled : DC – ≥100 MHz
Input Impedance, DC-Coupled : 1 M ±1% in parallel with 20 pF ±2.0 pF
Input Voltage, Maximum : ±300 V (DC or AC) CAT II; derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz and above
Time base system
Accuracy, Long Term Sample Rate and Delay Time : ±100 ppm over any ≥1 ms interval
Triggering system
Accuracy, Trigger Level, DC Coupled : ±(3% of |Setting – Net Offset1| + 0.2 div × volts/div setting + Offset Accuracy
Sensitivity, Edge-Type Trigger DC Coupled :
CH1 or CH2 0.35 division from DC to 20 MHz, increasing to 1 div at 100 MHz
External : 50 mV from DC to 20 MHz, increasing to 150 mV at 100 MHz
External/10 : 500 mV from DC to 20 MHz, increasing to 1.5 V at 100 MHz
Input Impedance, External Trigger : 1 M ±2% in parallel with 20 pF ±2 pF
Power requirements
Power Consumption : ≤65 Watts (120 VA)
Signal acquisition system
Accuracy, DC Gain, Envelope Acquisition Mode : ±2% for sec/div settings from 10 s/div to 5 ns/div
Accuracy, DC Voltage Measurement Sample Acquisition Mode : DC accuracy ±(2.0% × (|reading – Net Offset1|) + Offset
Accuracy + 0.13 div + 0.6 mV)