Objects Tiiips Categories
YOKOGAWA - DL7100 500MHz 4 tracce
"Descrizione"
by Al222 (20707 pt)
2018-May-21 18:30

Number of input channels 4 (CH1 to CH4)

Input coupling settings AC 1 MΩ, DC 1 MΩ, DC 50 Ω, GND

Input connector BNC

Input impedance 1 MΩ ±1.0%, approx. 20 pF

50 Ω ±1.0% (V SWR 1.4 or less (DC to 500 MHz))

Voltage-axis sensitivity setting 1 MΩ input : 2 mV/div to 10 V/div (1 - 2 - 5 steps)

50 Ω input : 2 mV/div to 1 V/div (1 - 2 - 5 steps)

Maximum input voltage 1 MΩ input (at 1 kHz or less) : 400 V (DC + AC peak) (282 Vrms CATII)

50 Ω input : 5 Vrms or less and 10 Vpeak or less

DC offset range (max) 2 mV/div to 50 mV/div : ±1 V

(At 1 : 1 probe attenuation) 100 mV/div to 500 mV/div : ±10 V

1 V/div to 10 V/div : ±100 V

Vertical (voltage) axis precision

DC precision*1 : ±(1.5% of 8 div + offset voltage precision)

Offset axis precision*1 2 mV/div to 50 mV/div : ±(1% of set value + 0.2 mV)

100 mV/div to 500 mV/div : ±(1% of set value + 2 mV)

1 V/div to 10 V/div : ±(1% of set value + 20 mV)

Frequency characteristics*1*2 50 Ω input

(–3dB point when sine wave 1 V/div to 10 mV/div : DC to 500 MHz

of amplitude ±4 div is input) 5 mV/div to 2 mV/div : DC to 400 MHz

1 MΩ input (defined at the tip of the probe when using the passive probe 700988)

10 V/div to 10 mV/div : DC to 400 MHz

5 mV/div to 2 mV/div : DC to 300 MHz

–3dB point for AC coupling used Not above 10 Hz (When using the 10 : 1 probe (standard accessory), 1 Hz or less)

Interchannel skew above 1 ns

(with identical settings)

Residual noise*3 Larger of ±1.25 mV or ±0.15 div (typical*4)

Interchannel isolation –34 dB (typical*4)

(at identical voltage sensitivity, DC to 500 MHz)

A/D conversion resolution 8 bits (24 LSB/div)

Probe attenuation settings 1 : 1, 10 : 1, 100 : 1, 1000 : 1

Bandwidth 100 MHz or 20 MHz band limit ON/OFF

Maximum sample rate Realtime sampling mode

With interleave ON : 1 GS/s

With interleave OFF : 500 MS/s

Repetitive sampling mode : 100 GS/s

Maximum record length 701410 With interleave ON : 2 MWords/CH

With interleave OFF: 1 MWords/CH

701420 With interleave ON : 8 MWords/CH

With interleave OFF: 4 MWords/CH

701430 With interleave ON : 4 MWords/CH

With interleave OFF: 2 MWords/CH

701440 With interleave ON : 16 MWords/CH

With interleave OFF: 8 MWords/CH

_______________________________________________________

*1 As measured following calibration (after 30-minute warmup), with internal-clock timebase, under standard operating conditions as described on page 16 - 10.

*2 For repetitive events

The frequency region for single shot is DC to sampling frequency/2.5 or the frequency region for the repetitive event, whichever is less.

*3 Measured under following conditions: input block shorted; 10 kWord record length; Normal acquisition mode; accumulation OFF; 1 : 1 probe attenuation

*4 Typical (or average) value; not guaranteed

Evaluate