HP 5384A Frequency counter
"Descrizione"
by CPU1 (1962 pt)
2021-Nov-14 16:34

Review Consensus: 10 Rating: 10 Number of users: 1
Evaluation  N. ExpertsEvaluation  N. Experts
1
  
6
  
2
  
7
  
3
  
8
  
4
  
9
  
5
  
10
  1

HP 5384A Frequency counter

The Keysight 5384A Frequency Counter utilizes interpolar-enhanced reciprocal counting to provide outstanding performance at a low cost. Frequency coverage allows measurements from 10 Hz to 225 MHz, with resolution of 9 digits per second. The 5384A incorporates a programmable alphanumeric display, accessible remotely via either HP-IB (standard) or HP-IL (optional). The liquid crystal display (LCD) reduces component count, is easy to read (indoors or out), and improves counter reliability and serviceability.

The half-rack width 5384A instrument is portable, rackable, and stackable for use in the field, at a workbench, or as part of an ATE system. Options include a four-hour-capacity battery pack, and a high-stability oven oscillator for higher accuracy.

Channel A Characteristics

  • Range: 10 Hz to 100 MHz
  • Sensitivity: [MAIN LEVEL] off

15 mVrms sine wave 50 Hz to 100 MHz

25 mVrms sine wave 10 Hz to 50 Hz

45 mV pk-pk 5 ns minimum pulse width

  • Dynamic Range: 45 mV to 4 V pk-pk × attenuator setting.
  • Coupling: AC
  • Impedance:

X1: 1 Mohm NOMINAL | | <25 pF

X20: 500 kohm NOMINAL | | <25 pF

  • Attenuator: X1 or X20 NOMIONAL, X20 increases to X40 below 50 Hz
  • Low Pass Filter: 100 kHz NOMINAL 3 dB point

Channel B Characteristics

  • Range: 50 to 225 MHz
  • Sensitivity: 10 mVrms 50 to 200 MHz, 15 mVrms 200 to 225 MHz
  • Dynamic Range: 10 mV to 1 Vrms
  • Coupling: AC
  • Impedance: 50 ohm NOMINAL
  • Attenuator Level:
  • Manual: variable from X1 to X5 (0 to 14 dB) NOMINAL
  • Auto: AGC mode for improved noise suppression.
  • Damage Level: 350 VDC + 5 Vrms AC

Timebase

  • Frequency: 10 MHz
  • Aging Rate: <3 × 10e-7/mo.
  • Temperature: <5 × 10e-6, 0 degrees to 50 degrees C. Ref. to 25 degrees C.
  • Line Voltage: <1 × 10e-7 for ±10% variation.

Frequency A and B

  • Range Channel A: 10 Hz - 100 MHz
  • Range Channel B: 50 MHz - 225 MHz
  • LSD Displayed: 10 Hz to 1 nHz
  • LSD ((4 nsec) / (Gate Time)) × FREQ
  • Resolution: ±1 LSD ± ((1.4 × Trigger Error + 1 nsec rms) / (Gate Time)) × Freq
  • Accuracy: ± Resolution ± Time Base Error × Period


  • High input sensitivity extends applications
  • Extensive input conditioning improves accuracy
  • Variable display digits improve readability
  • External dc operation extends portability
  • Proven reliability
  • For complete details, click on the Data Sheet link.
  • High Input Sensitivity
  • At least 15 mV rms sensitivity reduces the need for preamps and improves low-level signal measurement accuracy.
  • Extensive Input Conditioning
  • A low pass filter, automatic or manual attenuator and trigger level control features provide versatile signal-to-noise conditioning for optimum accuracy.
  • Variable Display Digits
  • View only readable digits from stable signals by controlling the number of display digits. Use as few as 3 or as many as 11 digits.
  • External dc Operation

  • Wide range (9-15V) external dc input power allows use of sources such as 12V automobile power to keep the timebase warm, or charge the battery option while in transit to remote sites.